Oxygen

            Oxygen is determined by both positive and negative ion APIMS. The positive ion mode can be used with either corona discharge or 63Ni sources.  However, the 63Ni source must be used for negative ion mode.

Positive Ion Mode

             In bulk gases such as nitrogen, helium, argon, oxygen is determined by monitoring O2+ primarily formed by the reaction

O2 + N4+ → O2+ + 2N2

We collect our signal in ion counting mode so that the standard deviation of the signal is the square root of the number of counts.  In analog mode the standard deviation only can be computed be physically measuring it.

            We observe about 1000 counts per second in positive ion mode so for a one second integration time 1000 counts are observed.  The signal to noise is the square root of the number of counts or 33 in this case.  Defining the lower limit of detection as the concentration at which the S/N is 3, the lower limit of detection is 100 pptv.  With an integration time of 100 sec the S/N = 10.

            However, like most determinations the LLD is determined by fluctuations in the blank.  In this case the blank is due mostly to H2O∙N2+.  A convenient way to determine the blank is to insert a very good purifier in line which removes the O2 leaving only the signal for H2O∙N2+.   The lower limit of detection in counts  based on the 95 percentile confidence limit is computed from the relationship

 

Here σ is the standard deviation of the signal,  ns is the number of signal measurements and no is the number of blank measurements.  In the above example if ns = no = 1000 and σ = 33 the lower limit of detection is 4 counts or 4/1000 =0.004 ppbv or 4 pptv for an integration time of 1 second.  Actually the trend in the absolute sensitivity of the instruments limits the length of time the signal can be studied before instrument drift dominates the signal.

Negative ion Mode  

In negative ion mode oxygen is usually the most electroegative ion species present.  In this case the available thermalized electrons attach to oxygen

O2 + e → O2-

O2- forms complexes with several species including H2O and NH3.  The oxygen concentration can be computed from the sum of the intensity of the O2- peak and all its complexes. In systems where H2O is the primary impurity other than O2 the oxygen concentration is proportional to the ratio of the O2- peak intensity and the O2- ∙H2O peak intensity.  In these systems the oxygen concentration can be determined with precision down to 10 pptv and perhaps lower.

            In liquid ammonia oxygen can be determine by following the signals for O2-, O2- ∙H2O and O2- ∙NH3.  The lower limit of detection is 100 pptv which is higher than for nitrogen because the ammonia is diluted 10/1 with clean nitrogen to prevent damage to the APIMS pumps.  The LLD can be restored to that observed in nitrogen if the pumps were replaced with ones than can survive in pure ammonia.